3T-C: Siemens MAGNETOM Prisma
Description
The 3T-C system is a Siemens Prisma scanner, installed in 2016. This is an FDA-approved scanner and currently serves the MRI imaging needs for high-volume neuroimaging projects, including the ABCD project and the HCP Aging, Development, Psychosis and Baby projects.
Magnet Information
- Magnet Manufacturer: Siemens
- Field Strength: 3.0 Tesla
- Shielding: Active
- Bore Diameter: 920 mm
- Installation Year: 2016
- Field Homogeneity: 2.0 (400) ppm (mm)
- Total Mass: 10 tons
- Windings Length: None km
- Windings Material:
- Helium Bath Volume: 1319 liters
- Stored Energy: None MJ
Console
- Console: Siemens MAGNETOM Prisma
- Software Platform: VE11C
RF Transmit and Receive Information
- Number of Receive Channels: 64
- Number of Transmit Channels: 2
- Transmit Power (per channel): None kW
- Combined Mode Transmit Power: 43 kW
- X-Nuclei Transmit Power: 8 kW
Primary Gradient
- Gradient Model:
- Maximum Strength: 80 mT/m
- Slew Rate: 200 T/m/s
- Rise Time: None µs
- 2nd Order Shims: True (10.0 A/ch)
- 3rd Order Shims: False
Clinical & Human Subjects
- Patient Accessible Bore Size: 600 mm
- Bore Length: 2150 mm
- Clinical Scans: No
- Human Subjects: Yes
Intercoms
- Avotec Silent Scan Intercom
- Siemens Intercom
Acquisition Peripherals
- FIRMM research
- Not Applicable MR Technologist Service Available
Stimulus computers
- Mac Stimulus Computer
- Matlab PsychToolbox
- Neurobs Presentation License (installed)
- PST E-Prime 2.0.10.353
- PsychoPy 1.83
- PsychoPy 2023
- Windows Matrix Control Computer
- Windows Stimulus Computer
Displays
- NEC Projector
Matrix Switch
- Extron 8x8 DVI Matrix Switch
Auditory
Response pads
Eyetrackers and cameras
- MREyes Eye Monitoring Camera